Risperidone and Metabolite
Risperdal Level
Test Codes
EPIC: LAB851, Beaker: XRISP, MAYO: FRISP
Department
Send Outs
Specimen Collection Criteria
Collect (preferred specimen): One dark Green-top sodium heparin tube.
Also acceptable: One plain Red-top tube.
Do not use Serum Separator Tubes.
Physician Office/Draw Specimen Preparation
Centrifuge to separate serum or plasma from cells within two hours of collection. Transfer serum or plasma to a plastic transport tube and refrigerate (2-8°C or 36-46°F).
Preparation for Courier Transport
Transport: 2.0 mL serum or plasma, refrigerated (2-8°C or 36-46°F). (Minimum: 0.6 mL)
Rejection Criteria
- Serum Separator (SST) tubes.
- Specimens not collected and processed as indicated.
In-Lab Processing
Centrifuge to separate serum or plasma from cells within two hours of collection. Transfer serum or plasma to a plastic transport tube and refrigerate (2-8°C or 36-46°F).
Transport: 2.0 mL serum or plasma, refrigerated (2-8°C or 36-46°F). (Minimum: 0.6 mL)
Storage
Specimen Stability for Testing:
Room Temperature (20-26°C or 68-78.8°F): 72 hours
Refrigerated (2-8°C or 36-46°F): 7 days
Frozen (-20°C/-4°F or below): 180 days
Specimen Storage in Department Prior to Disposal:
Specimen retention time is determined by the policy of the reference laboratory. Contact the Send Outs Laboratory with any questions.
Laboratory
Sent to Mayo Clinic Laboratories in Rochester, MN. Forwarded to Medtox Laboratories.
Performed
Monday – Sunday.
Results available in 3-7 days.
Reference Range
By report.
Test Methodology
Quantitative Liquid Chromatography – Tandem Mass Spectrometry (LC-MS/MS).
CPT Codes
80342
Contacts
Send Outs Laboratory – RO
248-551-9045
Name: Send Outs Laboratory – RO
Location:
Phone: 248-551-9045
Last Updated
11/18/2024
Microtainer® and Vacutainer® are registered trademarks of Becton, Dickinson and Company.
UroVysion® is a registered trademark of Abbott Laboratories. ThinPrep® is a registered trademark of Hologic, Incorporated.