Lead, Whole Blood (Southeast Michigan Only)
Please see Heavy Metals Screen, 24-Hour Urine
Test Codes
EPIC: LAB1231350, Beaker: XPBW, Warde: LEAD
Department
Send Outs
Instructions
The MDCH Blood Lead Analysis Report Form must be filled out and submitted with the specimen.
Specimen Collection Criteria
Collect (preferred specimen): One Royal Blue-top EDTA tube. (Minimum: 0.2 mL)
Also acceptable: One Lavender-top EDTA tube.
Note: Both venous and capillary blood draws are acceptable for this test.
Physician Office/Draw Specimen Preparation
Do not centrifuge. Refrigerate (2-8°C or 36-46°F) specimen prior to transport. Submit MDCH Blood Lead Analysis Report Form with the specimen.
Preparation for Courier Transport
Transport: 1.0 mL whole blood, refrigerated (2-8°C or 36-46°F). (Minimum: 0.3 mL)
Rejection Criteria
- Clotted specimens.
- Frozen specimens.
- Specimens not collected and processed as indicated.
In-Lab Processing
Do not centrifuge. Refrigerate (2-8°C or 36-46°F) specimen prior to transport. Submit MDCH Blood Lead Analysis Report Form with the specimen.
Transport: 1.0 mL whole blood, refrigerated (20-26°C or 68-78.8°F). (Minimum: 0.3 mL)
Storage
Specimen Stability for Testing:
Room Temperature (20-26°C or 68-78.8°F): Unacceptable
Refrigerated (2-8°C or 36-46°F): 1 month
Frozen (-20°C/-4°F or below): Unacceptable
Specimen Storage in Department Prior to Disposal:
Specimen retention time is determined by the policy of the reference laboratory. Contact the Send Outs Laboratory with any questions.
Laboratory
Sent to Warde Medical Laboratory, Ann Arbor, MI.
Performed
Monday – Friday.
Results available in 2-3 days.
Reference Range
By report.
Test Methodology
Atomic Absorption Spectrophotometry.
Interpretation
Elevated results may be due to skin or collection-related contamination, including use of a non-certified lead-free tube. Elevated levels of blood lead should be confirmed with a second specimen collected in a lead-free tube.
Clinical Utility
This assay aids in the evaluation of lead toxicity and poisoning.
CPT Codes
83655
Contacts
Send Outs Laboratory – RO
248-551-9045
Name: Send Outs Laboratory – RO
Location:
Phone: 248-551-9045
Last Updated
11/4/2024