IgE Soybean
Soybean (f14)
Test Codes
EPIC: LAB2111640, SOFT: ESO, Antrim #31389
Department
Immunology
Specimen Collection Criteria
Collect: One Gold-top SST tube.
Twenty individual allergen assays or allergen screens can be performed on one 5 mL Gold-top SST tube. Each allergen assay requires 100 mcL of serum.
Physician Office/Draw Specimen Preparation
Let specimen clot 30-60 minutes then immediately centrifuge to separate serum from cells. Refrigerate (2-8°C or 36-46°F) the centrifuged collection tube within two hours of collection.
Preparation for Courier Transport
Transport: Centrifuged collection tube, refrigerated (2-8°C or 36-46°F).
Rejection Criteria
- Plasma specimens.
- Severely lipemic or hemolyzed specimens.
In-Lab Processing
Let specimen clot 30-60 minutes then immediately centrifuge to separate serum from cells. Room temperature is acceptable for a maximum of two hours.
Storage
Specimen Stability for Testing:
Centrifuged SST Tubes and Microtainers® with Separator Gel
Room Temperature (20-26°C or 68-78.8°F): 2 hours
Refrigerated (2-8°C or 36-46°F): 7 days
Frozen (-20°C/-4°F or below): Unacceptable
Red-top Tubes and Microtainers® without Separator Gel
Room Temperature (20-26°C or 68-78.8°F): 2 hours
Refrigerated (2-8°C or 36-46°F): Unacceptable
Frozen (-20°C/-4°F or below): Unacceptable
Serum Specimens (Pour-Overs)
Room Temperature (20-26°C or 68-78.8°F): 2 hours
Refrigerated (2-8°C or 36-46°F): 7 days
Frozen (-20°C/-4°F or below): 1 month
Specimen Storage in Department Prior to Disposal:
Refrigerated (2-8°C or 36-46°F): 7 days
Laboratory
Royal Oak Special Chemistry Laboratory
Performed
Monday – Friday.
Results available the next business day.
Reference Range
Allergy Reference Range: Less than 0.35 kU/L.
Range (kU/L) |
Class |
Interpretation |
Less than or equal to 0.34 |
0 |
Negative |
0.35-0.69 |
1 |
Low |
0.70-3.49 |
2 |
Medium |
3.50-17.49 |
3 |
High |
17.50-49.99 |
4 |
Very High |
50.0-100.0 |
5 |
Very High |
Greater than 100 |
6 |
Very High |
Test Methodology
Fluorescence Enzyme Immunoassay (FEIA).
Interpretation
The allergen class may not be predictive of clinical disease in some patients. The diagnosis of allergy should be based upon patient history and clinical findings. The diagnosis of allergy should not be based upon laboratory findings alone.
Clinical Utility
Positive assay results indicate a high probability of allergic disease. Negative assay results effectively rule out allergy induced by those allergens.
Clinical Disease
Soybean flour and meal are commonly found in animal feed, infant formula, bakery goods, Chinese cooking, cereals, fillers in meat and candy products, and in certain topical dermatologic products (1).
The primary soybean allergens include 2S, 7S, 11S, and 15S. Cross-reactivity has been demonstrated between soybean allergens. Soybean sensitive persons usually do not produce an allergic reaction to the ingestion of soybean oil, probably because the soy proteins have been removed (2). Patients with soybean allergies are often sensitive to peanut allergens (3).
It is estimated that 25% of milk-sensitive persons will become allergic to soy protein after prolonged exposure to soybean allergens (3).
Reference
- Patterson, Roy, M.D., C. Raymond Zeiss, Jr., M.D., Leslie Carroll Grammar, M.D., Paul A. Greenberger, M.D. Allergic Diseases: Diagnosis and Management, 4th ed. J.B. Lippincott Company: Philadelphia, 1993, pg. 144.
- Metcalfe, Dean D. M.D., Hugh A. Sampson, M.D., Ronald A. Simon, M.D., Food Allergy: Adverse Reactions to Foods and Food Additives. Blackwell Scientific Publications: Boston, 1991, pg. 42-43, 346-347.
- Korenblat, Phillip E., M.D. & H. James Wedner, M.D. Allergy: Theory and Practice, 2nd ed. W.B. Saunders Company: Philadelphia, 1992, pg. 518-519.
CPT Codes
86003
LOINC: 6248-9
Contacts
Special Chemistry Laboratory – RO
248-551-8071
Name: Special Chemistry Laboratory – RO
Location:
Phone: 248-551-8071
Last Updated
11/20/2024
Microtainer® and Vacutainer® are registered trademarks of Becton, Dickinson and Company.
UroVysion® is a registered trademark of Abbott Laboratories. ThinPrep® is a registered trademark of Hologic, Incorporated.