Lab Test

Perampanel (Southeast Michigan Only)

Fycompa

Test Codes

EPIC: LAB1231458, Beaker: XPERAM, MAYO: FPERP

Department

Send Outs

Specimen Collection Criteria

Collect (preferred specimen): One plain Red-top tube.

Do not use Serum Separator Tubes.

Physician Office/Draw Specimen Preparation

Centrifuge to separate serum from cells within 2 hours of collection. Transfer serum to a plastic transport tube and refrigerate (2-8°C or 36-46°F).

Preparation for Courier Transport

Transport: 1.0 mL serum, refrigerated (2-8°C or 36-46°F). (Minimum: 0.6 mL)

Rejection Criteria

  • Serum Separator (SST) tubes.
  • Specimens not collected and processed as indicated.

In-Lab Processing

Centrifuge to separate serum from cells within 2 hours of collection. Transfer serum to a plastic transport tube and refrigerate (2-8°C or 36-46°F).

Transport: 1.0 mL serum, refrigerated (2-8°C or 36-46°F). (Minimum: 0.6 mL)

Storage

Specimen Stability for Testing:

Room Temperature (20-26°C or 68-78.8°F): 30 days
Refrigerated (2-8°C or 36-46°F): 30 days
Frozen (-20°C/-4°F or below): 1 year

Specimen Storage in Department Prior to Disposal:

Specimen retention time is determined by the policy of the reference laboratory. Contact the Send Outs Laboratory with any questions.

Laboratory

Sent to Mayo Clinic Laboratories in Rochester, MN. Forwarded to NMS Laboratories.

Performed

Monday – Sunday.
Results available in 8-12 days.

Reference Range

By report.

Test Methodology

Quantitative Liquid Chromatography/Tandem Mass Spectrometry (LC-MS/MS).

CPT Codes

80339

Contacts

Last Updated

11/11/2024

Microtainer® and Vacutainer® are registered trademarks of Becton, Dickinson and Company.
UroVysion® is a registered trademark of Abbott Laboratories. ThinPrep® is a registered trademark of Hologic, Incorporated.