Estradiol, Blood Level
E2, 17 ß-Estradiol, Estradiol Level, Estradiol (E2)
Test Codes
EPIC: LAB523, Antrim #17015
Department
Chemistry
Specimen Collection Criteria
Collect: One Gold-top SST tube. (Minimum Whole Blood: 4.0 mL)
Contact the Laboratory for acceptability of other tube types.
Physician Office/Draw Specimen Preparation
Let SST specimen clot 30-60 minutes then immediately centrifuge to separate serum from cells. Refrigerate (2-8°C or 36-46°F) the centrifuged SST tube within two hours of collection.
Preparation for Courier Transport
Transport: Centrifuged SST tube, refrigerated (2-8°C or 36-46°F). (Minimum: 1.0 mL)
Rejection Criteria
Red-top tubes with serum not separated from cells within two hours of collection.
In-Lab Processing
Let SST specimens clot 30-60 minutes. Centrifuge SST tubes or Microtainers® to separate serum from cells. Deliver immediately to the appropriate testing station.
Storage
Specimen Stability for Testing:
Centrifuged SST Tubes and Microtainers® with Separator Gel
Room Temperature (20-26°C or 68-78.8°F): 2-4 hours
Refrigerated (2-8°C or 36-46°F): 7 days
Frozen (-20°C/-4°F or below): Unacceptable
Red-top Tubes and Microtainers® without Separator Gel
Room Temperature (20-25°C or 68-77°F): 2-4 hours
Refrigerated (2-8°C or 36-46°F): Unacceptable
Frozen (-20°C/-4°F or below): Unacceptable
Serum Specimens (Pour-Overs)
Room Temperature (20-26°C or 68-78.8°F): 2-4 hours
Refrigerated (2-8°C or 36-46°F): 7 days
Frozen (-20°C/-4°F or below): 7 days
Specimen Storage in Department Prior to Disposal:
Refrigerated (2-8°C or 36-46°F): 7 days
Laboratory
Dearborn Chemistry Laboratory
Royal Oak Automated Chemistry Laboratory
Performed
Sunday – Saturday, 24 hours a day.
STAT results available within 1 hour of receipt to the Royal Oak Laboratory. (Outreach Only)
Results available within 24 hours of receipt in the Laboratory.
Reference Range
Males: 11-44 pg/mL.
Females:
Follicular: 21-251 pg/mL.
Mid cycle: 38-649 pg/mL.
Luteal: 21-312 pg/mL.
Post-menopausal: ≤28 pg/mL.
Test Methodology
Chemiluminescence Immunoassay.
Interpretation
- High Estradiol serum values are seen with: Ovarian tumors, adrenal feminizing tumors, precocious puberty (female), liver disease, and male gynecomastia.
- Low Estradiol serum values are seen with: Ovarian failure and oral contraceptives.
- CAUTION: Interference by Fulvestrant
NOTE: This assay should NOT be used to assess estradiol levels for patients undergoing Fulvestrant treatment. An alternate method, such as chromatography, should be used. All analogues of the estradiol molecule have the potential to cause interference/cross reactivity with the Abbott ARCHITECT Estradiol assay. If estradiol testing is required on such a patient, please order Estrogens, Serum Fractionated. The sample will then be sent to Mayo Medical Labs for testing by LC-MS.
Clinical Utility
Measurement of estradiol is used clinically in the investigation and management of fertility disorders, gynecomastia, estrogen-producing ovarian and testicular tumors and in hyperplasia of the adrenal cortex. Serum E2 is measured to determine the estrogen status of women, such as in some cases of amenorrhea, and as a guide to monitoring follicular development during ovulation induction and to avoid hyper-stimulation.
CPT Codes
82670
LOINC: 2243-4, 35384-7
Contacts
Chemistry Laboratory – DBN
313-436-2196
Name: Chemistry Laboratory – DBN
Location:
Phone: 313-436-2196
Automated Chemistry Laboratory – RO
248-551-8065
Name: Automated Chemistry Laboratory – RO
Location:
Phone: 248-551-8065
Last Updated
10/24/2024